MRI compatibility of position-sensitive photomultiplier depth-of-interaction PET detectors modules for in-line multimodality preclinical studies Articles uri icon

publication date

  • February 2013

start page

  • 83

end page

  • 87

volume

  • 702

International Standard Serial Number (ISSN)

  • 0168-9002

Electronic International Standard Serial Number (EISSN)

  • 1872-9576

abstract

  • This work addresses the feasibility of a small-animal, in-line PET/MR system based on Position-Sensitive Photo Multiplier Tubes (PS-PMTs). To this end, we measured the effects of static magnetic fields on the PS-PMTs performance in order to explore the minimal tandem separation between the PET and MR subsystems to preserve their respective performances. We concluded that it is possible to achieve minimal degradation of the PET scanner performance (after a system recalibration) if the magnetic field strength influencing the PET detectors is less than 1 mT and if it is oriented perpendicularly to the longitudinal axis of the tube. Therefore, we predict that it will be possible to maintain the PET image quality if it is placed outside the 1 mT line.

subjects

  • Biology and Biomedicine

keywords

  • pet/mri; photomultiplier; magnetic field; preclinical imaging; multimodality imaging; molecular imaging